GE RAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
8-144 Section 8-46 - Replacement of the T-DPI
Section 8-46
Replacement of the T-DPI
8-46-1 Manpower
One person, 30 minutes
8-46-2 Tools
Philips screwdriver, Allen Wrench, Hirose ejector
8-46-3 Preparations
1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4.
2.) Remove the REAR CABINET BEZEL ASSY BT16 and disconnect cables as described in 8-28-4
"Removal Procedure" 2), 3), 4) and 7) on page 8-93.
3.) Remove the NEST EMI BRKT BT16 as described in 8-56-4 "Removal Procedure"
4.) Remove 3 RS Knobs.