GE RAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
4-42 Section 4-8 - Mode Transition
Section 4-8
Mode Transition
4-8-1 B Mode
1.) Make sure that the probe is connected properly.
2.) Select the probe on the screen menu.
3.) Press [B] key on the operator panel to enter only B mode.
4.) Verify that B-mode scan starts.
4-8-2 M Mode
1.) Make shre that the probe is connected properly.
2.) Select the probe on the screen menu.
3.) Press [M] key on the operator panel.
4.) Verify that M-mode scan starts.
4-8-3 B/CF Mode
1.) Make sure that the probe is connected properly.
2.) Selec the probe on the screen menu.
3.) Press [B] key on the operator panel to enter only B mode.
4.) Press [CF] key on the operator panel.
5.) Verify that B/CF onde scan starts.
4-8-4 PW Mode
1.) Make sure that the probe is connected properly.
2.) Select the probe on the screen menu.
3.) Press [PW] key on the operator panel.
4.) Verify that PW mode scan starts.
4-8-5 CW (Optional)
NOTE: This test is only applied if CWD hardware option has been installed.
1.) Make sure that the probe, which supports CWD, is connected properly.
2.) Select the probe and select [CW] on the operator panel or on the application menu.
3.) Verify that CW mode scan starts.
4-8-6 3D/4D (Optional)
1.) Prepare a 3D/4D probe that the version of software has supported.
2.) Make sure that the probe is connected properly.
3.) Select the probe on the operator panel.
4.) Press 3D or 3D/4D key
5.) Verify that Static 3D image is acquired.