GE DRAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-181
3.) Open keyboard assembly.
4.) Remove ECG DUMMY BRKT.
Figure 8-255 ECG DUMMY BRKT
8-57-5 Installation Procedure
1.) Assemble ECG kit on keyboard base.
2.) Screw ECG kit with Keyboard base using 2 screws (2159625-screw w/SP PH m4x8)and fix ECG
USB cable by using the hook of bracket.
Figure 8-256 ECG kit Installation