GE DRAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-41
Section 7-9
User Interface Troubleshooting
7-9-1 Console Test
1.) Connect the USB "standard GE" service dongle on the front USB connector(s). As soon as the
dongle is detected the login window pops up.
2.) Enter the "Standard GE Revolving" password (password changes every six months) and then click
ENABLE
.
3.) Start a probe.
4.) Press the F2
key on the control panel.
5.) In the "Utilities" menu, select SYSTEM SETUP
to invoke the setup desktop on the screen.
6.) On the left side of the screen select ADMINISTRATION
and then click the SERVICE tab.
7.) In the displayed window, select the SERVICE TOOLS
button. The "Service Tools" window appears
on the screen; see: Figure 7-45 on page 7-47.
8.) Click the CONSOLE TEST
button.
Figure 7-36 Sevice Tools