GE RAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
8-156 Section 8-52 - Replacing optional Peripherals
Section 8-52
Replacing optional Peripherals
8-52-1 Manpower
One person, 70 minutes
8-52-2 Tools
Standard Field Service Tool Kit
NOTICE
!! NOTICE:
Normally auxiliary devices and peripherals come pre-installed with the Voluson™ P8 / Voluson™ P6
system.
NOTICE
!! NOTICE:
READ and UNDERSTAND these instructions thoroughly before proceeding with the installation.
Perform each step in sequence and check it off when completed. If a problem occurs after completion,
go back through the procedure and check for implementation errors before contacting your Ultrasound
Region.
CAUTION
!! CAUTION:
Possible Operational damage. Failure to strictly follow ESD (Electrostatic Discharge)
precautions during this installation may cause constant or intermittent operational
abnormalities. Strictly follow all precautions.