EasyManua.ls Logo

GE Voluson P8 BT16 - 7-7-3-3 DRFM Overall Test

GE Voluson P8 BT16
612 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE RAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
7-32 Section 7-7 - Hardware Tests
7-7-3-3 DRFM Overall Test
This diagnostic test performs the self-diagnosis of each electric component and interface in the DRFM
board.
Table 7-2 below outlines Voluson™ P8 / Voluson™ P6 Diagnostics Tool described in the sub-sections.
7-7-3-3-1 Voltage DRFM
It verifies Each Power Voltage Level on DRFM board, and displays the current voltage levels and the
test result as PASSED/FAILED.
7-7-3-3-2 Temperature DRFM
It verifies each GPU, CPU, DPS temperature measure on DRFM and displays the current temperature
status and the test result as PASSED/FAILED.
7-7-3-3-3 IIC DRFM Interface
It verifies I2C path named I2C_Local by using EEPROM on DRFM board and displays the test result as
PASSED/FAILED.
7-7-3-3-4 HW ID Config
It verifies HW ID configuration on DRFM board and displays the current HW ID and the test result as
PASSED/FAILED.
7-7-3-3-5 FPGA DRFM Interface
It verifies FPGA(DRFM) configuration on DRFM board by reading/writing test data and displays the test
result as PASSED/FAILED.
Table 7-2 Automatic diagnostic test under DRFM Test
Sub-section Description Page Number
7-7-3-3-1
Voltage DRFM
7-38
7-7-3-3-2
Temperature DRFM
7-38
7-7-3-3-3
IIC DRFM Interface
7-38
7-7-3-3-4
HW ID Config
7-38
7-7-3-3-5
FPGA DRFM Interface
7-38

Table of Contents