GE DRAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5-57
5-15-1-3 Rollback
This function offers the possibility to simply restore the previously used system configuration (rollback),
which was stored on R:\ during “FMI from DVD”.
1.) Turn system OFF and then back ON.
2.) As soon as the “Boot Screen” appears (see: Figure 5-31 on page 5-56), press the PG DN
[ Arrow down] key on the keyboard until the Rollback item is highlighted, then press ENTER.
3.) When the following WARNING message appears, press the
Arrow left] button to highlight OK
and then press ENTER
.
After performing the rollback, the system reboots.
The Ultrasound Application is started and finally the 2D screen is displayed on the monitor.
After 3 sec. without pressing any key, the system will boot-up in windows (= Voluson item).
If you missed step 2, retry again with CTRL + ALT + DEL.
Figure 5-32 Warning message