GE DRAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-9
2.) On the network sniffer screen select Capture -> Options.
3.) Choose your Network in the “Interface” pull-down menu (1) -> Intel Network or Wireless adapter.
NOTE: Remember, if the device has a DVR, it will show up in the “Interface” list.
NOTE: If you are working with a Wireless connection, you must uncheck the “Capture packets in promiscuous
mode” (A). The wireless option does not support this mode.
4.) If desired, select a “Capture Filter” from the pull-down menu (2); see
5.) Adjust “Display Options”(3) for best performance to avoid packet loss on high throughput.
6.) Select Start(4).
7.) Minimize network sniffer window and initiate communication to the DICOM device (e.g., send
Figure 7-8 Capture -> Options
Figure 7-9 Capture Options

Table of Contents