GE RAFT VOLUSON™ P8 / VOLUSON™ P6
DIRECTION 5723243, R
EVISION 6 DRAFT (MAY 23, 2018) BASIC SERVICE MANUAL
7-24 Section 7-6 - How to use the Auto Tester program
11.)Start scanning. You can scan normally and everything will be recorded to the program (up to several
hours.)
NOTE: It is important that you are recording the processes where the errors normally occur.
The following screen will appear.
12.)Select the “Save” icon. The file will be saved in D:Export\AutotesterScripts\*.*.
13.)Press Alt + D to export system data and the needed logs. Operation see
NOTE: he standalone recorded "Auto Tester" file makes only sense with the Alt + D log! To analyze the
workflow the exact date and time of occurrence needs to be documented!
14.)If desired, save the script file also on DVD/CD. Therefore:
- Insert an empty DVD/CD+R/RW in the Drive and select the “Burn on CD” icon.
- Enter a Filename.
- After the DVD/CD write is finished click the OK button.
15.)Close the “Auto Tester” program.
Stop the program by clicking on [Stop] shown on the screen,
or by pressing the ALT GR
key on the alphanumeric keyboard.
Figure 7-27 Auto Tester Finished