EasyManua.ls Logo

Helmer iBX080 - Sensor Calibration; View Sensor Calibration Values

Helmer iBX080
73 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
360179-A/B 23
Installation & Conguration
8.3.1.1 Temperature Setpoint
NOTE Default chamber temperature setpoint is 4.0 ºC.
Setpoint can be changed from the i.C
3
and from the Haemonetics
®
BloodTrack
®
interface.
The temperature setpoint is the desired temperature of the HaemoBank™ for stored product.
Change the temperature setpoint if:
► Your organization requires a chamber temperature other than 4.0 ºC
Perform the following:
1 Touch Home, i.C³ APPS, Settings.
2 Enter the Settings password.
3 Touch + or on the Temperature Setpoint spin box to select desired setting.
4 Touch Home.
8.4 Sensor Calibration
Sensor calibration values are programmed at the factory. Calibration values can be viewed through the
i.C³ Sensor Calibration screen.
Sensor Calibration screens
8.4.1 View Sensor Calibration Values
1 Touch Home, i.C³ APPS, Settings.
2 Enter the Settings password.
3 Touch Sensor Calibration.
► Sensor o󰀨set values and their current temperature readings are displayed
4 Touch Home.
Setting Initial Factory Value
Simulated product
probe
Varies (set at the factory to match
a NIST traceable independent
thermometer)
Control probe
Air probe
Compressor probe

Table of Contents

Related product manuals