360179-A/B 35
Maintenance & Service
9.6.8 i.C³
®
Touchscreen
NOTICE The i.C³ touchscreen is sensitive to excessive pressure when touched. Use light
pressure when cleaning the touchscreen.
Required tools:
► Soft cotton cloth
Clean the screen:
1 Using a soft cotton cloth, wipe the screen clean.
9.6.9 BloodTrack
®
Kiosk
NOTICE The BloodTrack
®
kiosk is sensitive to excessive pressure when touched. Use light
pressure when cleaning the touchscreen.
Required tools:
► Soft cotton cloth
Clean the screen:
1 Using a soft cotton cloth, wipe the screen clean.
9.7 Bypass System Check and Reset
NOTICE ► Please refer to the BloodTrack Courier
®
User Guide for details on preparing the
HaemoBank™ refrigerator for service. This will include instruction on how to open
the main door.
► If trays are not latching or releasing properly, rotate the Bypass Release handle
between the locked and unlocked position. If this does not release or correct the
problem, an inspection of the latching mechanism will be required and must be done
from the rear of the unit. Contact Haemonetics
®
Corporation BloodTrack
®
Customer
Support for further assistance.
1 Insert key into Bypass Release handle and unlock handle.
2 Rotate the Bypass Release handle 90 degrees, unlatching the compartments.
3 Verify all trays are unlatched by pulling the trays approximately half way out of the compartment.
Leave trays in this position.
4 If all trays are unlatched, secure the Bypass Release handle by rotating 90 degrees back to the
locked position.
5 Push each tray back into position and ensure they cannot be pulled back out.
6 Listen for the lock to engage.
7 From the BloodTrack
®
interface, conrm return to normal operation and ensure the Bypass Alarm has
cleared.
NOTICE Refer to the BloodTrack
Courier
®
User Guide for details on returning the HaemoBank™
refrigerator to normal use.