360179-A/B 28
Maintenance & Service
9.1.3 Power Failure Alarm Test
1 Change the Power Failure delay setting to 0 minutes.
a Touch Home, i.C³ APPS, Settings.
b Enter the Settings password.
c Touch Alarm Settings.
d Touch on the Power Failure spin box to change the value to 0.
2 Switch the AC ON/OFF switch OFF.
► Power Failure alarm will activate immediately
3 Switch the AC ON/OFF switch ON.
► Power Failure alarm will clear and audible alarm will cease
4 Change the Power Failure time delay to the original setting.
a Touch Home, i.C³ APPS, Settings.
b Enter the Settings password.
c Touch Alarm Settings.
d Touch + on the Power Failure spin box to change the value to the original setting.
5 Touch Home.
9.1.4 Door Open Alarm Test
1 Change the Door Open (Time) delay setting to 0 minutes.
a Touch Home, i.C³ APPS, Settings.
b Enter the Settings password.
c Touch Alarm Settings.
d Touch on the Door Open (Time) spin box to change the value to 0.
2 Open the door.
► Door Open alarm will activate immediately
3 Close the door.
► Door Open alarm will clear and audible alarm will cease
4 Change the Door Open (Time) setting to the original setting.
a Touch Home, i.C³ APPS, Settings.
b Enter the Settings password.
c Touch Alarm Settings.
d Touch + on the Door Open (Time) spin box to change the value to the original setting.
5 Touch Home.
9.2 Upgrade System Firmware
Helmer may occasionally issue updates for the i.C³ rmware. Follow the upgrade instructions included
with the rmware update.
9.3 Test and Replace Backup Batteries
9.3.1 Test the i.C
3
®
Monitoring System / Access Control Backup Battery
Required tools:
► 4’ ladder or step stool