6.6 Connect AC Power Cord for Compartmental Access Refrigerator
NOTE The Detachable AC Power Cord for the refrigerator is shipped in the accessory box
included in the top of the Compartmental Access Refrigerator.
1 Connect the Compartmental Access Refrigerator AC power cord to the receptacle on top of the refrigerator.
2 Connect the Compartmental Access Refrigerator AC power cord to a grounded outlet meeting the
electrical requirements stated in Section I, Item 4.
6.7 Connect Backup Power
6.7.1 Power-up i.C
3
Backup Battery
The i.C
3
Monitoring System and Access Control magnetic lock have a backup battery system, enabling a
period of continuous operation if power is lost.
The backup battery ON/OFF switch is located on the top of the Compartmental Access Refrigerator.
i.C
3
Monitoring System / Access Control backup battery ON/OFF switch (circled)
The backup battery is switched OFF for shipping. Switch the backup battery ON/OFF switch ON to
provide backup power to the monitoring system and Access Control magnetic lock.
NOTE The i.C
3
Monitoring System and Access Control magnetic lock will start on battery power
alone. If the Compartmental Access Refrigerator is not connected to AC power and the
battery is switched on, the i.C
3
Monitoring System and Access Control magnetic lock will
begin running on battery power.
Battery life varies by manufacturer as well as voltage level remaining. If full battery power is available and
no battery-related alarms are active, backup power for the i.C
3
Monitoring System and Access Control
magnetic lock is available for up to 20 minutes.
The Low Battery alarm will activate when battery power is almost depleted.
NOTICE During a power failure:
The backup battery does not provide continued refrigeration of the chamber. The
chamber temperature may rise above the established limits necessary to maintain
integrity of stored product.
The backup battery provides power to the monitoring system and Access Control
magnetic lock until battery power is depleted.
The backup battery is rapidly depleted by the Access Control magnetic lock.
To preserve battery life during extended power failure, switch the backup battery ON/
OFF switch OFF and use the door key to manually secure the exterior door.