360179-A/B 24
Installation & Conguration
8.4.2 Change Sensor Calibration Oset Values
NOTE ► Contact Haemonetics
Corporation BloodTrack
Customer Support for instructions on
changing sensor calibration oset values.
8.5 Factory Default Settings
Factory settings may be simultaneously returned to factory default values.
NOTE ► The Factory Default Settings may not be the same as the settings that were factory-
calibrated before the Compartmental Access Refrigerator was shipped.
► If Factory Default Settings are restored, all calibration values and User PINs will be
lost.
General Settings Restored Value
Home Screen Application Icons i.C³ APPS, Settings, Temperature Graph, Automatic
Alarm Test, Information Logs
Temperature Setpoint 4.0 ºC
Password (for Settings screen) 1234
Password Protection (for Settings screen) On
Sounds On
Alarm Volume 9
Alarm Tone 2
Unit ID No default value
Date Format MM/DD/YY
Day Not aected (maintained in real-time clock)
Month
Year
Time Format 12-hour
Minute Not aected (maintained in real-time clock)
Hour
AM/PM
Language English
Temperature Units ºC
Temperature Graph Screensaver O
Alternate Screensaver On
Min/Max Temperature Display On
Air Probe Temperature Display O
Air Probe Temperature Alarm O
Display Brightness High (3 symbols)
Temperature Calibration Settings Values entered at the factory
Access Control Touchpad On
Alarm Settings Restored Value
Upper Probe High Temperature (Setpoint) 5.5 ºC
Upper Probe High Temperature (Time Delay) 0 minutes