360179-A/B 51
Troubleshooting
Problem Possible Cause Action
The Low/No Battery
alarm is activating
sporadically.
Battery voltage level
on the rechargeable
backup battery for the
monitoring system is
low
Ensure battery has had 24 hours to recharge after
an extended use.
Replace the backup battery for the monitoring
system.
11.7 Testing Problems
Problem Possible Cause Action
The automatic
temperature tests do
not work.
High Alarm setpoint
is set signicantly
higher than the default
value, or the Low
Alarm setpoint is set
signicantly lower
than the default value
Conrm the alarm setpoints are set at the
appropriate values.
Test the temperature alarms manually.
Connections for
the upper chamber
temperature probe are
loose
Test the probe connections. Secure the
connections if necessary.
Upper chamber
temperature probe is
faulty
Conrm the probe is reading correctly. Calibrate
the probe if necessary.
Conrm the probe is providing resistance in the
range of 98 Ω to 110 Ω. Replace the probe if
necessary.
Control board is faulty
Replace parts with those included in the control
board kit, or replace the monitor/control board.
11.8 Condensation Problems
Problem Possible Cause Action
There is excessive
water in the water
evaporation tray.
Heater in the
condensate
evaporator tray is
faulty
Conrm the heater is hot and is drawing the
appropriate current.
For 115 V Compartmental Access Refrigerator, the
current should be approximately 0.43 A to 0.55 A.
For 230 V Compartmental Access Refrigerator, the
current should be approximately 0.21 A to 0.27 A.
Condensate
evaporator tray fan is
faulty
Conrm both fans are running.
Contact Haemonetics
Corporation BloodTrack
Customer Support.
Humid air is entering
the chamber
Conrm the Compartmental Access Refrigerator
is level, and the doors are aligned, closing
tightly, and sealing correctly. Correct issues as
necessary.
Ensure cabinet penetrations are sealed.
Condensate
evaporator tray fan
blower is faulty
Conrm fan is running.
Contact Haemonetics
Corporation BloodTrack
Customer Support.