360179-A/B 50
Troubleshooting
Problem Possible Cause Action
The Compartmental
Access Refrigerator is
connected to power,
but the AC Power
Failure alarm is
active.
Circuit breaker is
tripped or faulty
Reset or replace the circuit breaker.
AC ON/OFF power
switch is OFF
Turn the ON/OFF AC power switch to the ON
position.
AC ON/OFF power
switch is faulty
Replace the ON/OFF AC power switch.
Power cord is faulty
Conrm the power cord is connected securely.
Secure the power cord if necessary. Replace
power cord if necessary.
Outlet connection is
faulty
Verify power at the outlet. Repair the original
outlet or connect to a di󰀨erent outlet if necessary.
A component is
faulty or internal
connections are loose
Contact Haemonetics
Corporation BloodTrack
Customer Support.
The Door Open
alarm is activating
sporadically.
Doors are not closing
completely
Conrm hinges are free of dirt or debris. Clean if
necessary.
Conrm the hinge cams are not damaged.
Replace the cams if necessary.
Conrm rear door is locked.
Connections for the
door switch are faulty
Test the switch connections. Secure the
connections if necessary.
Ensure door sensor bracket pushes switch closed.
One or both door
switches are faulty
Replace the door switch or switches.
Compartmental
Control PCB
assembly is faulty
Contact Haemonetics
Corporation BloodTrack
Customer Support.
The Condenser alarm
is active.
Condenser alarm
setpoint is too low
Conrm the alarm setpoint is at the appropriate
value (refer to Section II, Item 8.3).
Compressor is
overheating due to a
lack of air ow
Check the condenser grill and clean if necessary
(refer to Section III, Item 9.6.1).
Conrm the Compartmental Access Refrigerator
is correctly located (refer to Section II, Item 5 for
location requirements).
Condenser fan motor
is faulty
Replace the condenser fan motor.
Connections for
the condenser
temperature probe are
loose.
Test the probe connections. Secure the
connections at CP board if necessary.
Condenser
temperature probe is
faulty
Test the probe. Replace the probe if necessary.
Conrm the probe is providing resistance in the
range of 78 Ω to 120 Ω.
A component is
faulty or internal
connections are loose
Contact Haemonetics
Corporation BloodTrack
Customer Support.