360179-A/B 25
Maintenance & Service
Alarm Settings Restored Value
Upper Probe Low Temperature (Setpoint) 2.0 ºC
Upper Probe Low Temperature (Time Delay) 0 minutes
Air Probe High Temperature (Setpoint) 14.0 ºC
Air Probe High Temperature (Time Delay) 3 minutes
Air Probe Low Temperature (Setpoint) -6.0 ºC
Air Probe Low Temperature (Time Delay) 3 minutes
Compressor Temperature (Setpoint) 50.0 ºC
Compressor Temperature (Time Delay) 5 minutes
Power Failure 1 minute
Probe Failure 0 minutes
Door Open (Time) 10 minutes
8.6 Restore Factory Default Settings
NOTICE ► If Factory Default Settings are restored, all calibration values and User PINs will be
lost.
► Contact Haemonetics
Corporation BloodTrack
Customer Support prior to restoring
Factory Default Settings.
Restore settings:
1 Touch Home, i.C³ APPS, Settings.
2 Enter the Settings password.
3 Touch Restore Factory Settings.
► The “Are you sure you want to restore factory settings?” message is displayed
► Touch √. The message screen closes and factory settings are restored
► Touch X. The message screen closes and factory settings are not restored
4 Touch Home.
8.7 View System Information
1 Touch Home, i.C³ APPS, Contacts.
► The Contacts screen is displayed
2 Touch More Info.
► The Versions screen is displayed, including software conguration, software revision numbers,
hardware serial numbers, i.C³ IP address, and i.C³ MAC address
3 Touch Home.