360179-A/B 25
Maintenance & Service
Alarm Settings Restored Value
Upper Probe Low Temperature (Setpoint) 2.0 ºC
Upper Probe Low Temperature (Time Delay) 0 minutes
Air Probe High Temperature (Setpoint) 14.0 ºC
Air Probe High Temperature (Time Delay) 3 minutes
Air Probe Low Temperature (Setpoint) -6.0 ºC
Air Probe Low Temperature (Time Delay) 3 minutes
Compressor Temperature (Setpoint) 50.0 ºC
Compressor Temperature (Time Delay) 5 minutes
Power Failure 1 minute
Probe Failure 0 minutes
Door Open (Time) 10 minutes
8.6 Restore Factory Default Settings
NOTICE If Factory Default Settings are restored, all calibration values and User PINs will be
lost.
Contact Haemonetics
Corporation BloodTrack
Customer Support prior to restoring
Factory Default Settings.
Restore settings:
1 Touch Home, i.C³ APPS, Settings.
2 Enter the Settings password.
3 Touch Restore Factory Settings.
The “Are you sure you want to restore factory settings?” message is displayed
► Touch . The message screen closes and factory settings are restored
► Touch X. The message screen closes and factory settings are not restored
4 Touch Home.
8.7 View System Information
1 Touch Home, i.C³ APPS, Contacts.
► The Contacts screen is displayed
2 Touch More Info.
► The Versions screen is displayed, including software conguration, software revision numbers,
hardware serial numbers, i.C³ IP address, and i.C³ MAC address
3 Touch Home.