360179-A/B 34
Maintenance & Service
Clean painted interior surfaces:
1 Apply mild detergent solution to the painted interior surfaces.
2 Using a cotton cloth, wipe the interior surfaces dry.
Clean stainless steel interior surfaces:
1 Apply general-purpose laboratory cleaner to the stainless steel interior surfaces.
2 Using a cotton cloth, wipe the interior surfaces dry.
9.6.5 Door Gaskets
NOTE Avoid getting water in the Access Control magnetic lock.
Required tools:
► Soft cotton cloth (2)
► Mild soap and water solution
Clean the door gaskets:
1 Using a soft cotton cloth, apply soap and water solution to the door gaskets and gasket crevices.
2 Using a second clean and dry cotton cloth, wipe the door gaskets clean.
9.6.6 Clean the Tray Assembly
NOTE ► Refer to Section III, Item 10.3.1, for instructions on removing the tray from the
compartment assembly.
Required supplies:
► Soft cotton cloth
► Mild detergent solution
Clean trays and covers:
1 Apply mild detergent solution to the tray and cover.
2 Using a cotton cloth, wipe the tray and cover dry.
9.6.7 Clean Probe Bottle and Rell Solution
Required supplies:
► Fresh water-bleach solution
► 1:10 ratio of bleach to water
► Bleach is 5% solution of commercial sodium hypochlorite (NaOCl) (Equivalent oxidizing cleaner/
disinfectant approved by your organization may be substituted)
► 4 oz. (120 mL) of product simulation solution per bottle
► 10:1 ratio of water to glycerin
Clean and rell bottle:
1 Remove probe(s) from bottle.
2 Remove bottle from bracket.
3 Clean bottle with water-bleach solution.
4 Rinse bottle thoroughly with tap water.
5 Fill bottle with 4 oz. (120 mL) of product simulation solution.
6 Cap bottle tightly to minimize evaporation.
7 Place bottle in bracket.
8 Replace probe(s), immersing at least 2” (50 mm).