360179-A/B 34
Maintenance & Service
Clean painted interior surfaces:
1 Apply mild detergent solution to the painted interior surfaces.
2 Using a cotton cloth, wipe the interior surfaces dry.
Clean stainless steel interior surfaces:
1 Apply general-purpose laboratory cleaner to the stainless steel interior surfaces.
2 Using a cotton cloth, wipe the interior surfaces dry.
9.6.5 Door Gaskets
NOTE Avoid getting water in the Access Control magnetic lock.
Required tools:
► Soft cotton cloth (2)
► Mild soap and water solution
Clean the door gaskets:
1 Using a soft cotton cloth, apply soap and water solution to the door gaskets and gasket crevices.
2 Using a second clean and dry cotton cloth, wipe the door gaskets clean.
9.6.6 Clean the Tray Assembly
NOTE Refer to Section III, Item 10.3.1, for instructions on removing the tray from the
compartment assembly.
Required supplies:
► Soft cotton cloth
► Mild detergent solution
Clean trays and covers:
1 Apply mild detergent solution to the tray and cover.
2 Using a cotton cloth, wipe the tray and cover dry.
9.6.7 Clean Probe Bottle and Rell Solution
Required supplies:
► Fresh water-bleach solution
► 1:10 ratio of bleach to water
► Bleach is 5% solution of commercial sodium hypochlorite (NaOCl) (Equivalent oxidizing cleaner/
disinfectant approved by your organization may be substituted)
► 4 oz. (120 mL) of product simulation solution per bottle
► 10:1 ratio of water to glycerin
Clean and rell bottle:
1 Remove probe(s) from bottle.
2 Remove bottle from bracket.
3 Clean bottle with water-bleach solution.
4 Rinse bottle thoroughly with tap water.
5 Fill bottle with 4 oz. (120 mL) of product simulation solution.
6 Cap bottle tightly to minimize evaporation.
7 Place bottle in bracket.
8 Replace probe(s), immersing at least 2” (50 mm).