Loading...
11.1 Connecting the D/A Output Terminal
157
1
2
3
4
5
6
7
8
9
10
11
When the D/A output function is used, measurement data such as
the voltage, current, and power is output in analog (D/A) form.
Using D/A Output
(3169-21 only)
11
11.1 Connecting the D/A Output Terminal
To avoid electrocution, turn off the power to all devices
before pluggingor unplugging any of the D/A output connec-
tors.
To prevent damage to the instrument, never connect or discon-
nect the connector with the power on (the D/A output is insulated
from the voltage input and current input).
Four output channels are available. Exercise great care when
handling, as these channels are not insulated from each other.
Use the optional 9441 connection cable.
To prevent damage to the 3169-20/21, do not short-circuit the
output terminal or input a voltage.