11
1
2
3
4
5
6
7
8
9
10
11
To avoid damage to the product, do not short-circuit the output
terminal and do not input voltage to the output terminal.
Voltage input terminals U
1
, U
2
, and U
3
are common to the N ter-
minal and are not insulated. To avoid the risk of electric shock,
do not touch the terminals.
Note that the product may be damaged if current or voltage
exceeding the selected measurement range is applied for a long
time
When the power is turned off, do not apply voltage or current to
the voltage input terminals or clamp sensor. Doing so may dam-
age the product.
To prevent damage to the instrument and sensor, never connect
or disconnect a sensor while the power is on, or while the sensor
is clamped around a conductor.
When the voltage or current for the power line being measured
exceeds the maximum input for this device, use an external
VT(PT) or CT.
When using an external VT(PT) or CT, make sure you use a
device with a minimal phase difference. By setting the VT(PT)
or CT ratio, you can read measurement values directly.
Using VT(PT) and CT