13.2 Range Configuration and Accuracy by Clamp-On-Sensor
187
13
2
3
4
5
6
7
8
9
10
11
Power Range Configuration
(when the 9694 or 9695-02 is used)
Accuracy by Clamp-On Sensor
(when the 9694 or 9695-02 is used)
Voltage Wiring
Current
9695-02 CLAMP ON SENSOR (CAT III 300 V)
9694 CLAMP ON SENSOR (CAT III 300 V)
500.00 mA 1.0000 A 5.0000 A 100.00 A 500.00 A
150.00 V 1P2W 75.000 W 150.00 W 750.00 W 15.000 kW 75.000 kW
1P3W
3P3W2M
3P3W3M
150.00 W 300.00 W 1.5000 kW 30.000 kW 150.00 kW
3P4W
3P4W4I
225.00 W 450.00 W 2.2500 kW 45.000 kW 225.00 kW
300.00 V 1P2W 150.00 W 300.00 W 1.5000 kW 30.000 kW 150.00 kW
1P3W
3P3W2M
3P3W3M
300.00 W 600.00 W 3.0000 kW 60.000 kW 300.00 kW
3P4W
3P4W4I
450.00 W 900.00 W 4.5000 kW 90.000 kW 450.00 kW
600.00 V 1P2W 300.00 W 600.00 W 3.0000 kW 60.000 kW 300.00 kW
1P3W
3P3W2M
3P3W3M
600.00 W 1.2000 kW 6.0000 kW 120.00 kW 600.00 kW
3P4W
3P4W4I
900.00 W 1.8000 kW 9.0000 kW 180.00 kW 900.00 kW
• The range-configuration table shows the full-scale display
value of each measurement range.
• Voltage and current measurements are indicated as 0.4% to
130% f.s. of the range. If a measurement is below 0.4% f.s., it
will be zero-suppressed.
• Power measurement is indicated as 0% to 130% f.s. of the range.
It will be zero-suppressed when the voltage or current is 0.
• The range configuration for apparent power (S) and reactive
power (Q) is the same, except that the unit is changed to VA
and var, respectively.
• When the VT ratio and CT ratio are set, the ranges will be mul-
tiplied by (VT ratio x CT ratio) (when a range falls below 1.0000
mW or exceeds 9.9999 GW, a scaling error occurs and the set-
ting is not accepted).
• The accuracy-guarantee ranges of the 9694 and 9695-02 sen-
sors are 500 mA to 5 A, respectively.
Range 9694 CLAMP ON SENSOR 9695-02 CLAMP ON SENSOR
50.000 A --
±
0.5%rdg.
±
0.12%f.s.
10.000 A --
±
0.5%rdg.
±
0.2%f.s.
5.0000A
±
0.5%rdg.
±
0.12%f.s.
±
0.5%rdg.
±
0.3%f.s.
1.0000A
±
0.5%rdg.
±
0.2%f.s.
±
0.5%rdg.
±
1.1%f.s.
500.00mA
±
0.5%rdg.
±
0.3%f.s.
±
0.5%rdg.
±
2.1%f.s.