13.2 Range Configuration and Accuracy by Clamp-On-Sensor
Power Range Configuration
(when the 9667 is used)
Accuracy by Clamp-On Sensor
(when the 9667 is used)
Voltage Wiring
Current
9667 FLEXIBLE CLAMP ON SENSOR
500 A range 5000 A range
500.00 A 5.0000 kA
150.00 V 1P2W 75.000 kW 750.00 kW
1P3W
3P3W2M
3P3W3M
150.00 kW 1.5000 MW
3P4W
3P4W4I
225.00 kW 2.2500 MW
300.00 V 1P2W 150.00 kW 1.5000 MW
1P3W
3P3W2M
3P3W3M
300.00 kW 3.0000 MW
3P4W
3P4W4I
450.00 kW 4.5000 MW
600.00 V 1P2W 300.00 kW 3.0000 MW
1P3W
3P3W2M
3P3W3M
600.00 kW 6.0000 MW
3P4W
3P4W4I
900.00 kW 9.0000 MW
The range-configuration table shows the full-scale display
value of each measurement range.
Voltage and current measurements are indicated as 0.4% to
130% f.s. of the range. If a measurement is below 0.4% f.s., it
will be zero-suppressed.
Power measurement is indicated as 0% to 130% f.s. of the range.
It will be zero-suppressed when the voltage or current is 0.
The range configuration for apparent power (S) and reactive
power (Q) is the same, except that the unit is changed to VA
and var, respectively.
When the VT ratio and CT ratio are set, the ranges will be mul-
tiplied by (VT ratio x CT ratio) (when a range falls below 1.0000
mW or exceeds 9.9999 GW, a scaling error occurs and the set-
ting is not accepted).
Range 9667 FLEXIBLE CLAMP
ON SENSOR
5000 A range
9667 FLEXIBLE CLAMP
ON SENSOR
500 A range
5.0000 kA
±
2.2%rdg.
±
0.4%f.s.
500.00 A
±
2.2%rdg.
±
0.4%f.s.