13.2 Range Configuration and Accuracy by Clamp-On-Sensor
Power Range Configuration
(when the 9667 is used)
Accuracy by Clamp-On Sensor
(when the 9667 is used)
Voltage Wiring
Current
9667 FLEXIBLE CLAMP ON SENSOR
500 A range 5000 A range
500.00 A 5.0000 kA
150.00 V 1P2W 75.000 kW 750.00 kW
1P3W
3P3W2M
3P3W3M
150.00 kW 1.5000 MW
3P4W
3P4W4I
225.00 kW 2.2500 MW
300.00 V 1P2W 150.00 kW 1.5000 MW
1P3W
3P3W2M
3P3W3M
300.00 kW 3.0000 MW
3P4W
3P4W4I
450.00 kW 4.5000 MW
600.00 V 1P2W 300.00 kW 3.0000 MW
1P3W
3P3W2M
3P3W3M
600.00 kW 6.0000 MW
3P4W
3P4W4I
900.00 kW 9.0000 MW
• The range-configuration table shows the full-scale display
value of each measurement range.
• Voltage and current measurements are indicated as 0.4% to
130% f.s. of the range. If a measurement is below 0.4% f.s., it
will be zero-suppressed.
• Power measurement is indicated as 0% to 130% f.s. of the range.
It will be zero-suppressed when the voltage or current is 0.
• The range configuration for apparent power (S) and reactive
power (Q) is the same, except that the unit is changed to VA
and var, respectively.
• When the VT ratio and CT ratio are set, the ranges will be mul-
tiplied by (VT ratio x CT ratio) (when a range falls below 1.0000
mW or exceeds 9.9999 GW, a scaling error occurs and the set-
ting is not accepted).
Range 9667 FLEXIBLE CLAMP
ON SENSOR
5000 A range
9667 FLEXIBLE CLAMP
ON SENSOR
500 A range
5.0000 kA
±
2.2%rdg.
±
0.4%f.s.
500.00 A
±
2.2%rdg.
±
0.4%f.s.