13.2 Range Configuration and Accuracy by Clamp-On-Sensor
185
13
2
3
4
5
6
7
8
9
10
11
Power Range Configuration
(when the 9669 is used)
Accuracy by Clamp-On Sensor
(when the 9669 is used)
Voltage Wiring
Current
9669 CLAMP ON SENSOR
100.00 A 200.00 A 1.0000 kA
150.00 V 1P2W 15.000 kW 30.000 kW 150.00 kW
1P3W
3P3W2M
3P3W3M
30.000 kW 60.000 kW 300.00 kW
3P4W
3P4W4I
45.000 kW 90.000 kW 450.00 kW
300.00 V 1P2W 30.000 kW 60.000 kW 300.00 kW
1P3W
3P3W2M
3P3W3M
60.000 kW 120.00 kW 600.00 kW
3P4W
3P4W4I
90.000 kW 180.00 kW 900.00 kW
600.00 V 1P2W 60.000 kW 120.00 kW 600.00 kW
1P3W
3P3W2M
3P3W3M
120.00 kW 240.00 kW 1.2000 MW
3P4W
3P4W4I
180.00 kW 360.00 kW 1.8000 MW
• The range-configuration table shows the full-scale display
value of each measurement range.
• Voltage and current measurements are indicated as 0.4% to
130% f.s. of the range. If a measurement is below 0.4% f.s., it
will be zero-suppressed.
• Power measurement is indicated as 0% to 130% f.s. of the range.
It will be zero-suppressed when the voltage or current is 0.
• The range configuration for apparent power (
S
) and reactive
power (
Q
) is the same, except that the unit is changed to VA
and var, respectively.
• When the VT ratio and CT ratio are set, the ranges will be mul-
tiplied by (VT ratio x CT ratio) (when a range falls below 1.0000
mW or exceeds 9.9999 GW, a scaling error occurs and the set-
ting is not accepted).
Range 9669 CLAMP ON SENSOR
1.0000 kA
±
1.2%rdg.
±
0.11%f.s.
200.00 A
±
1.2%rdg.
±
0.15%f.s.
100.00 A
±
1.2%rdg.
±
0.2%f.s.