Notes on Use
────────────────────────────────────────────────────
NOTE
・ Always set the power supply frequency before measurement. When it has
been set incorrectly, inaccurate measurement would be performed. How to set
this, Section 3.3, "Selecting the Power Supply Frequency."
・ Accurate measurement may be impossible in locations subject to strong
external magnetic fields, such as transformers and high-current conductors, or
in locations subject to strong external electric fields, such as radio
transmission equipment.
・ For accurate measurement, allow the unit to warm up for 30 minutes before
starting the operation.
・ In the source of current (SOURCE), there is a fuse for the circuit protection.
When it is ruptured, the measurement should be impossible.
・ Do not use such relay as dealing with the small signals, for fear of breaking
the pellicle of the contact point.
・ When the one that includes the L component such as the transformer for the
power supply a lot is measured a measured value may not stabilize.