Notes on Use
────────────────────────────────────────────────────
NOTE
Always set the power supply frequency before measurement. When it has
been set incorrectly, inaccurate measurement would be performed. How to set
this, Section 3.3, "Selecting the Power Supply Frequency."
Accurate measurement may be impossible in locations subject to strong
external magnetic fields, such as transformers and high-current conductors, or
in locations subject to strong external electric fields, such as radio
transmission equipment.
For accurate measurement, allow the unit to warm up for 30 minutes before
starting the operation.
In the source of current (SOURCE), there is a fuse for the circuit protection.
When it is ruptured, the measurement should be impossible.
Do not use such relay as dealing with the small signals, for fear of breaking
the pellicle of the contact point.
When the one that includes the L component such as the transformer for the
power supply a lot is measured a measured value may not stabilize.