EasyManua.ls Logo

Hioki 3227 - Page 110

Hioki 3227
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
98
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Synta
x
:FREQuency data
dat
a
50, 60 (numerical data in NR1 format)
Functio
n
Sets the use power supply frequency.
Exampl
e
Transmission
:FREQ 50
The use power supply frequency is set to 50 Hz.
Synta
x
:FREQuency?
Functio
n
Returns the present use power supply frequency as a numerical data in
NR1 format (data).
Respons
e
If headers are ON
If headers are OFF
:FREQUENCY data
data
Exampl
e
If headers are ON
If headers are OFF
Transmission
Response
:FREQ?
:FREQUENCY 50
:FREQ?
50
Synta
x
:FUNCtion data
dat
a
RESIstance, TEMPerature (character data)
Functio
n
Sets the measurement mode either the resistance measurement or the
temperature measurement.
Exampl
e
Transmission
:FUNC RESI
The measurement mode is set to the resistance mode.
:FREQuency
:FREQuency?
:FUNCtion

Table of Contents

Related product manuals