122
────────────────────────────────────────────────────
8.9 Sample Programs
────────────────────────────────────────────────────
NOTE
(5) Setting and executing the temperature correction
Summary This program sets and executes the temperature correction.
This program sets the standard temperature and the temperature coefficient
of the temperature correction.
When the value impossible to set is input, an error is generated, and the
setting keeps as it is.
Program List
100 OUTPUT 701;":FUNC RESI"
110 LINPUT "Reference? (‑10.0 99.9) ",R$
120 LINPUT "Parameter? (‑9999 9999) ",P$
130 OUTPUT 701;":TCSET "&R$&"."&P$
140 OUTPUT 701;":TC ON"
150 END
Comments
100 Set the measurement function to the resistance measurement mode
110 Designate the standard temperature
120 Designate the temperature coefficient
130 Set the standard temperature and the temperature coefficient
140 Execute the temperature correction