3.2 Turning on the Power Supply
────────────────────────────────────────────────────
When the temperature correction
function is selected
When the temperature conversion
function is selected
NOTE
Display the name of this unit, the version number and the power supply
frequency settings.
TC on the display indicates the function to be used, the temperature correction
function or temperature conversion function. When TC on the display is
blinking, the temperature conversion function is used.
(7) Enter the normal measurement state.
・ Before measurement, perform the warm up at least for 30 minutes after
turning on the power supply.
・ This unit contains the back up function. The previous settings appear, when
turning on the power supply. The back up function works until the internal
lithium battery has consumed.
・ The battery would run down after four years. Please contact your nearest
service representative for the battery change.
・ It is not possible to use the temperature correction function and the
temperature conversion function at the same time.
・ In resistance measurement mode, if pressing the
TC key switches to the ℃
indication on the display, the temperature conversion function can be used.