6.3 Electrical Specifications
────────────────────────────────────────────────────
COMP
―――――――
Comparator table or
temperature conversion
table
COMP
―――――――
Comparator table or
temperature conversion
table
3 2 1 0 3 2 1 0
H H H H No change L H H H 8
H H H L 1 L H H L 9
H H L H 2 L H L H 10
H H L L 3 L H L L 11
H L H H 4 L L H H 12
H L H L 5 L L H L 13
H L L H 6 L L L H 14
H L L L 7 L L L L 15
Selecting the comparator table or the temperature conversion table (COMP 0
―――――――――
to COMP 3
―――――――――
)
During in the temperature conversion display, these signals selects the
temperature conversion tables.
Select the comparator table or the temperature convesion table followingly
during the resistance measurement and execute.
See Section 5.2, "Executing the Comparator", for executing the comparator.
Measurement trigger (TRIG
――――――
)
When change this signal from "H" to "L" during holding the measurement
(trailing edge), the measurement and the displays are renewed.
See Section 4.6, "Holding the Measurement Value", for the holding the
measurement.
Requirement for the comparator output (MANU
――――――
)
When the external control terminal mode of the comparator settings is EXT,
during this signal is "L", the comparator results are output.
See Section 5.1, "Setting the Comparator", for setting the comparator.
Zero adjustment requirement (0 ADJ
――――――
)
When change this signal from "H" to "L" during the resistance measurement
(trailing edge), the zero adjustment is executed.
See Section 4.3.3, "Zero Adjustment", for the zero adjustment.
Print requirement (PRINT
――――――
)
This signal is used when printing by the 9203 DIGITAL PRINTER or the
centronics printer.
See Chapter 9, "Printer Interface", for the printer.