80
_____________________________________________________________________________________________
7.4 Command Reference
______________________________________________________________________________________________
Syntax
[:SENSe:]VOLTage[:DC]:REFerence?
[:SENSe:]VOLTage:AC:REFerence?
[:SENSe:]CURRent:DC:REFerence?
[:SENSe:]CURRent:AC:REFerence?
[:SENSe:]RESistance:REFerence?
[:SENSe:]LPResistance:REFerence?
[:SENSe:]FRESistance:REFerence?
[:SENSe:]LPFResistance:REFerence?
[:SENSe:]CLAMp:DC:REFerence?
[:SENSe:]CLAMp:AC:REFerence?
[:SENSe:]FREQuency:REFerence?
data
NR3 numerical data
Function
Inquires about the Zero-Adjusted value.
Response
syntax
Headers: ON :SENSE:VOLTAGE:AC:REFERENCE data
Headers: OFF data
Example
Headers: ON
Transmission CURR:DC:REF?
Response :SENSE:CURRENT:DC:REFERENCE 6.320000E-02
Headers: OFF
Transmission CURR:DC:REF?
Response 6.320000E-02
The Zero-Adjusted value in a DC current measurement is 63.2 mA.
Syntax
[:SENSe:]VOLTage[:DC]:REFerence:STATe data
[:SENSe:]VOLTage:AC:REFerenceSTATe data
[:SENSe:]CURRent:DC:REFerence:STATe data
[:SENSe:]CURRent:AC:REFerence:STATe data
[:SENSe:]RESistance:REFerence:STATe data
[:SENSe:]LPResistance:REFerence:STATe data
[:SENSe:]FRESistance:REFerence:STATe data
[:SENSe:]LPFResistance:REFerence:STATe data
[:SENSe:]CLAMp:DC:REFerence:STATe data
[:SENSe:]CLAMp:AC:REFerence:STATe data
[:SENSe:]FREQuency:REFerence:STATe data
data
ON/1: Zero-Adjust function enabled.
OFF/0: Zero-Adjust function disabled.
Function
Selects whether the Zero-Adjusted value set for the specified measurement function is to be
subtracted or not.
Example
Transmission :SENS:CURR:AC:REF:STAT 1
Enables the Zero-Adjust function of the AC current measurement function.
RES:REF:STAT OFF
Disables the Zero-Adjust function of the resistance measurement function.
Inquiry concerning the Zero-Adjusted value.
Setup for Zero-Adjust execution