EasyManua.ls Logo

Hioki 3237 - Page 96

Hioki 3237
169 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
84
_____________________________________________________________________________________________
7.4 Command Reference
______________________________________________________________________________________________
Syntax
:CALCulate:LIMit:BEEPer data
data
Character data
OFF,IN,FAIL,HL
Function
Sets the buzzer sound to beep according to the comparator determination.
Note
HL is the same as FAIL.
Example
Transmission :CALC:LIM:BEEP IN
Activates the buzzer when the comparator determination is IN.
Syntax
:CALCulate:LIMit:BEEPer?
data
Character data
OFF,IN,HL
Function
Inquires about the setting that activates the buzzer at the comparator determination.
Response
syntax
Headers: ON :CALCULATE:LIMIT:BEEPER data
Headers: OFF data
Example
Headers: ON
Transmission :CALC:LIM:BEEP?
Response :CALCULATE:LIMIT:BEEPER FAIL
Headers: OFF
Transmission :CALC:LIM:BEEP?
Response FAIL
If the determination is FAIL (Hi, Lo), the buzzer beeps.
Syntax
:CALCulate:LIMit:UPPer data
data
NRf numerical data
-199999 to 199999
Function
Sets the upper-limit value to be used for comparator determination in a count value.
Note
For example, if the upper-limit value is set to 123456 in the DC2 V range, the comparator
determines a Hi setting when the measurement value is greater than 1.23456 V.
Example
Transmission :CALC:LIM:UPP 156000
Sets the comparator upper-limit value to 156000.
Setup for comparator buzzer sound
Inquiry concerning comparator buzzer sound
Setup for comparator upper-limit value