87
────────────────────────────────────────────────────
8.1 Specifications
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
8.1.1 GP-IB Interface
NOTE
Interface Functions Provided
SH1
All source handshake functions
AH1
All acceptor handshake functions
T5
Basic talk functions, Serial poll function
Talk-only mode is provided.
The talker cancellation function with MLA (My Listen Address) is provided.
L4
Basic listener functions
Listen-only mode is not provided.
The listener cancellation function with MTA (My Talk Address) is provided.
SR1
All service request functions
RL1
All remote/local functions
PP0
Parallel polling is not provided.
DC1
All device clear functions
DT1
All device trigger functions
C0
The controller function is not provided.
ASCII codes are used.
Chapter 8
GP-IB/RS-232C Interface
8.1 Specifications
Compliance standard: IEEE-488.1 1987
Reference standard: IEEE-488.2 1987
On the 3332, if the output queue becomes full, it is cleared and a query error
is generated. This does not correspond to the clearing of the output queue
and the outputting of a query error in the deadlock state as stipulated in
IEEE 488.2. (A deadlock state occurs when both the input buffer and the
output queue are full, and processing cannot continue normally.)
This unit does not fully conform to IEEE488.2, and to the extent possible
should be used only for indicated reference data. Also, take care to ensure
that each command does not result in overflows of the input buffer and
output queue.