83
────────────────────────────────────────────────────
7.4 External Control Terminal
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
5 V (open)
0 V (short)
Integration start Integration stop
Integrate time
400 ms or more
Integration Start/Stop
NOTE
External control
5 V (open)
0 V (short)
Resets integration value
during this term
Over 300 ms
Reset Integration
NOTE
7.4 External Control Terminal
The external control terminals accept a 0/5 V logic signal or a relay contact
short/open signal to control the unit.
For details of the settings, refer to Sections 4.9, "Setting the Interface",
Section 4.3, "Setting the Integration", and Section 4.8, "Setting the Printer
Output".
(1) Integration start and stop (START/STOP terminal)
If integration is started by external control, the integration set time is ignored
and can only be stopped by external control. If it is attempted to stop
integration by pressing the
START/STOP
key, "Err.011" is displayed.
Integration's setting display will be the following way below.
(2) Reset integration
During integration operating (RUN indicator lit), "Err.015" is displayed and
it is not possible to reset integration.