19
────────────────────────────────────────────────────
2.5 Self Test
────────────────────────────────────────────────────
NOTE
2.5 Self Test
Indications in self test shifts as follows:
(1) All indicators illuminate
(2) Displays
a: Model
b: Software version number
(3) Setting conditions of the GP-IB address and RS-232C
(4) After self checking, error number is displayed if an unusual matter is
founded.
(5) Returns to normal measurement state.
・ For the error indications, refer to Section 6.3, "Error Indications."
・ Settings at powering on become previous condition. (Backup function)
However, this function does not work when internal lithium battery is used
up. In this case, the error is indicated after self-test.
・ The battery would run down after six years. Please contact your nearest
service representative for battery change.
・ Don't press the keys unnecessarily during self testing.
・ If the
SHIFT
keyispressedandthenthe
RESET
key is pressed during the
self-test, the unit will reset the system. (Refer to Section 6.2, "System
Resetting.")
・ If the measured line is active when the 3332 is tuned on, the 3332 may
malfunction or Err105 may be displayed. Thus, turn on the 3332 first and
check that Err105 is not displayed, and then turn on the measured line.