35
────────────────────────────────────────────────────
4.3 Setting the Integration
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
4.3.4 Notes on Integration
Ah or Wh flashes
(3) Integration operation
To start the integration after the previous integration value is reset:
Press the
SHIFT
RESET
key (RUN indicator off), and then press
START/STOP
key (RUN indicator is lit). The integration will start.
To add to the previous integration value.
Press the
START/STOP
key.
To stop forcibly the operating integration:
Press the
START/STOP
key. The RUN indicator is flashing and the
integration operation will stop.
To reset the integration value:
Press the
SHIFT
RESET
key, and the RUN indicator is turned off. The
integration value will be reset.
(4) Integration stop (RUN indicator flashing)
Reaches integrating time.
Press the
START/STOP
key while the RUN indicator is lit. (RUN indicator will
be flashing).
When integrate time reaches 10000 hours.
When integrate value reaches ±999999 MAh (MWh).
(5) Integration reset
Press the
SHIFT
RESET
key while the RUN indicator is flashing (RUN
indicator will be turned off).
(1) When starting integration, auto range setting of voltage and current is
canceled and the range at starting point is fixed. It is necessary to set the
range so that it will not be out of range during integrating.
If the input of current or active power for channels exceeds 130% of range
during integration, the value for the channel is integrated as input of
130%f.s.
The Ah or Wh unit indication flashes to indicate that an out-of-range value
has occurred. This continues flashing until the integration value is reset.