4.3 Setting the Integration
────────────────────────────────────────────────────
ENTER
Changing value
Moving digits
Display:
Supplement
Key Key operation
AVE・FREQ
PT・CT・SC
COMP1
COMP2
D/A
PRINT
RANGE SEL
INTERFACE
Moves to each setting screen.
4.3.3 Integration Operation
(2) Set the integration.
1. To select the integration time, press the or Setup key to change the
value.
Pressing or Setup key moves flashing digit.
2. The setting range of integrate time is 10 seconds to 10,000 hours. 
3. Press the
ENTER
key. The flashing will stop and the setting will be decided.
Unless this operation is done, the internal setting cannot be changed.
If the settings need to be changed again, press either the , , , Setup
keys and the flashing points will reappear.
If the integration time is set to 00000:00:00 or more than 10000 hours, enter
100000:00:00 by the
ENTER
key.
(3) Move to the measurement display.
Press the
INTEG
or
EXIT
key. The SHIFT indicator will turn off.
Other keys operations in Integration Setting mode
(SHIFT indicator is lit):
(1) Understanding the integrating state.
Integration execution state: RUN indicator lit.
Integration stopped state: RUN indicator flashing
Integration reset state: RUN indicator off.
(2) The limitations of this unit due to integration.
During integration execution state (RUN indicator lit) and when in integration
stopped state (RUN indicator flashing), voltage range, current range, rectifier
type, PT ratio, CT ratio, SC, and range setting cannot be changed.
("Err.012" or "Err.017" is displayed for a second.)