6.3 Error Indications
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
Error number
Error Indication List
Error
number
Errors
Pages
001
Backup error (*1, *2, *3)
Requires repair
101
ROM error (*3)
Requires repair
102
RAM error (*3)
Requires repair
103
RAM display (*3)
Requires repair
104
Timer circuit error (*3)
Requires repair
105
A/D and D/A converter error (*3)
Requires repair
011
(Integration) Key operation and external control used together
36,83
012
(Integration) Invalid key press during integration operation
22,25,34,36
013
(Integration) Integration started while scaling error has
36
014
(Integration) Restarted from maximum integration count or
occurred integration elapsed time
37
015
(Integration) Integration reset carried out during integration
36,83
016
Invalid key is pressed during display hold
21,22,25,27,29
017
Invalid key is pressed during setting
27,29,34,36
018
All range are set to OFF.
63
021
(Interface) Data out or help operation is carried out while it is
not in Pr. mode or when interface is not installed
185,193,194
022
(Interface) No space left in internal data buffer
186,189
023
(Interface) Invalid communication
37
*1 Err.001 (backup error)
Pressing any key frees the display and returns to the normal measurement state. The
3332, however, undergoes a system reset. (Refer to Section 6.2 "System Resetting".)
*2 Error.001 occurs when backup data stored in RAM is corrupted, which can result from
a depleted internal lithium backup battery, external noise ingress or loss of power while
changing settings. Please contact your nearest service representative for battery change.
The battery would run down after six years.
*3 Err.101 to 105, 001
Indicate internal problems with the 3332. If these error indications appear, the unit
requires repair.
NOTE
6.3 Error Indications
An error indicator appears when a malfunction is discovered during the
power-on self-test, or when an invalid key press occurs during integration or
display hold operations.
If the measured line is active when the 3332 is tuned on, the 3332 may malfunction or
Err105 may be displayed. Thus, turn on the 3332 first and check that Err105 is not
displayed, and then turn on the measured line.