27
────────────────────────────────────────────────────
3.5 Holding the Display
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
3.5.1 Restrictions During the Hold Status
Item Display
Selecting integrate time
Selecting number of averaging
Selecting frequency measurement source and range
Selecting D/A output items
Setting the PT ratio
Setting the CT ratio
Setting the SC
Selecting range
3.5 Holding the Display
Press the
HOLD
key, the HOLD indicator is lit and preserve all measuring
value at that point. (Hold state)
Press the
HOLD
key again, this indicator goes off, and return normal
measuring condition.
・Shifting of response, range and rectifier type are not allowed. ("Err.016" is
displayed.)
・Even if the auto-ranging is set for voltage, current, or frequency the range
cannot be changed. The range is fixed when the range is held.
If the HOLD state is released, auto-ranging function restarts.
・The following settings cannot be changed, which are explained in Chapter 4,
"Settings and Operations". ("Err.017" is displayed.)
Pressing the
HOLD
key to release the hold state (the HOLD indicator goes off)
changes setting.