36
────────────────────────────────────────────────────
4.3 Setting the Integration
────────────────────────────────────────────────────
Items
RUN indicator
Lit Flashing
Response
Rectifier type
Voltage and current range
Averaging time
Frequency range and source
PT ratio, CT ratio, and SC
D/A output item
Comparator
Range select
Integration time
Print time and item
Interface settings
●: changeable -: can not change
("Err.012" or "Err.017" is displayed for a second.)
(2) The following settings cannot be changed during integration (RUN indicator
lit) or integration stop (RUN indicator flashing), (Err.012 is displayed).
When integrated by external control, the displays will be as follows and the
setting cannot be changed.
To change the setting for the integration value reset, return the measurement
display with the
INTEG
key and press the
SHIFT
RESET
key.
(3) While the RUN indicator for integration is lit, resetting is out of order
(Err.015 is displayed). Resetting must be done after stopping integration
(interface and external control).
(4) On starting and stopping integration key operating (and :INTEG:STATE
START command of interface) and external control are unable to mix
("Err.011" is displayed).
(5) When using external control, the integrate time which is set according to
Section 4.3.2, "Integration Setting Method" is invalid. After all, integration
stops in following cases.
Stopped by external control signals.
When the integrate time achieves 10000 hours.
When the integrate value achieves ±999999 MAh (MWh).
(6) During displaying "S. Err", integrating cannot start ("Err.013" is displayed).
In this case, change PT/CT ratios, and SC. Refer to Section 4.7, "Setting the
PT ratio, CT ratio, and SC".