109
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Queries whether or not all action has been completed during execution.
Syntax
*OPC?
Function
The same as the *OPC command, except in that, at the instant that the
previous commands have been completed, instead of bit 0 (the OPC bit) of the
standard event status register (SESR) being set to 1, the response message "1"
is returned.
Notes
With this query, if any error occurs, no response message is produced.
No header is affixed to the response message.
Response
syntax
Headers: ON/OFF
"1"
Errors
Query error/ If the response message is longer than 1000 bytes.
System error/ This command is not executed. A device dependent error occurs.
Performs device initial setting.
Syntax
*RST
Function
Resets the 3332 unit. The parameters which are reset, the values to which they
are reset, and those items which are not affected by this command, are listed
below.
Parameters which are reset, and their new values:
Display V, A, W, PF
Response AUTO
Voltage range Auto ranging
Current range Auto ranging
Rectifier type RMS
Frequency source, range V, auto ranging
Integration time 10000:00:00 (10,000 hours)
PT/CT ratios, SC 1.000
D/A output item VA
Number of averaging 1
Comparator OFF (both CH1 and CH2)
Comparator upper limit 0.0000k (all tables)
Comparator lower limit 0.0000k (all tables)
Comparator items (CH1) W
Comparator items (CH2) A
Range selection All ON
HOLD OFF (continuous display)
Printing item V, A, W, +Wh, TIME
Printing interval 00:00:00 (interval off)
Response data header ON
Data separator semicolon ";"
*
OPC?
*
RST