EasyManua.ls Logo

Hioki 3332 - Page 174

Hioki 3332
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
160
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Queries the PT ratio.
Syntax
SCALe:PT?
Function
The current value of PT ratio is returned as a numerical value in NR2
numerical format.
Note
With this query, if any error occurs, no response message is produced.
Response
syntax
Headers: ON
Headers: OFF
":SCALE:PT_<NR2>"
"<NR2>"
Errors
Query error/ If the response message is longer than 1000 bytes.
System error/ This query is not executed. A device dependent error occurs.
Example
Transmission
Response
Headers: ON
Headers: OFF
":SCAL:PT?"
":SCALE:PT 5.000"
"5.000"
Sets the SC.
Syntax
SCALe:SC_<NR2>
<NR2>
0.001 to 9999.0
Function
The current value of SC is set.
<NR2> can be accepted in NRf format, but its effective value will be obtained
by rounding off the fifth digit on the basis of 5 and above being rounded up
and 4 and below being rounded down.
Errors
Execution error/ If the setting data is other than listed above.
Command error/ If the setting data is not in NRf format.
Device dependent error/
In the event of a system error
If this command is executed in the hold state
(HOLD indicator lit or flashing)
During integrating (RUN indicator lit)
Example
Transmission
":SCAL:SC 10.00"
The SC is set to 10.00.
SCALe:PT?
SCALe:SC

Table of Contents