165
────────────────────────────────────────────────────
8.4 Command Reference
────────────────────────────────────────────────────
Sets the voltage range.
Syntax
VOLTage:RANGe_<NR1>
<NR1>
15, 30, 60, 150, 300, 600
Function
Sets the voltage range. The units are volts (V).
<NR1> can be accepted in NRf format, but its effective value will be obtained
by rounding off the sixth digit on the basis of 5 and above being rounded up
and 4 and below being rounded down.
Notes
If the value specified is not a valid range setting, the next largest range is
selected.
A unit (V) should not be set to data portion.
After changing range, wait at least 5 seconds for the internal circuit to stabilize
before taking readings.
If the range which is set to OFF is selected, the range selection setting is
enabled.
Errors
Execution error/ If the setting data is other than listed above.
Command error/ If the setting data is not in NRf format.
Device dependent error/
In the event of a system error
If this command is executed in the hold state
(HOLD indicator lit or flashing)
During integrating (RUN indicator lit)
Example
Transmission
":VOLT:RANG 300.004"
The range is set to 300 V by rounding off.
Transmission
":VOLT:RANG 300.005"
The range is set to 600 V by rounding off.
VOLTage:RANGe