8.9 Notes on Interface
────────────────────────────────────────────────────
Symptom Cause / Treatment
The response message to a query
differs from the display on the
front panel of the 3332.
Due to the response message being produced at the instant
that the 3332 receives the query, there is a possibility that it
may not agree with the display at the instant that the
controller reads it in.
Although the talk only has been
set, printing has not work.
Has the printer been set to listen only?
Except 3332 and printer, has any other equipments of
controller or talk only been connected.
Although the printer mode has
been set, printing has not work.
Has 3332 been set to listen only ?
Has print interval time been set to 000:00:00? In this case,
set interval time.
Pressing DATA OUT key or HELP
key displays "Err 021".
Has GP-IB interface been equipped?
Has the 3332 been set to Pr. mode?