8
────────────────────────────────────────────────────
1.5 Display Reference
────────────────────────────────────────────────────
Error indications Meaning
Backup error (press any key to deactivate, needs repair)
ROM malfunction (cannot deactivate, needs repair)
RAM malfunction (cannot deactivate, needs repair)
RAM display malfunction (cannot deactivate, needs repair)
Timer circuit malfunction (cannot deactivate, needs repair)
A/D,D/A converter malfunction (cannot deactivate, needs repair)
*1(see below)
Mix up of control by key and remote during integration (displayed approx.
1 second)
A non-executable key was pressed during integration (RUN indicator
lit/flashing) (displayed approx. 1 second)
Integration was started while S.Err was displayed (displayed approx. 1
second)
Integration was started when it was ±999999M or 10000 hours
(displayed approx. 1 second)
Data was reset during integration (displayed approx. 1 second)
A invalid key was pressed during HOLD (HOLD indicator lit/flashing)
(displayed approx. 1 second)
A invalid key was pressed during display settings (displayed approx. 1
second)
All selected range was set to OFF.
Key was pressed when the printer mode is not set for the interface
(displayed approx. 1 second)
There is no room in the printer output memory (displayed approx. 1
second)
The interface cannot be communicated.
Other indications Meaning
In integration time setting, this indicates operation under external control.
Data is sent to printer after recovery from power failure.
Resets the 3332
*1
If the measured line is active when the 3332 is tuned on, the 3332 may
malfunction or Err105 may be displayed. Thus, turn on the 3332 first and
check that Err105 is not displayed, and then turn on the measured line.