212
────────────────────────────────────────────────────
11.3 Function Specifications
────────────────────────────────────────────────────
(20) Battery backup
Backup item
Setting values in each functions, integrating status and integration
value.
Backup time
6 years or more (23℃ (73
。
F) reference value)
(22) External interface
GP-IB INTERFACE
Based on IEEE-488.1 1987 and refer to IEEE-488.2 1987
Interface function: SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT1, C0
Remote control by controllers
Data output to the listen-only printer (manual, interval, help)
(PRINT lamp lit)
It is possible to set the interval with unit in 10 seconds from 10
seconds to 100 hours.
Remote release by LOCAL key (RMT lamp off)
RS-232C INTERFACE
Start-stop transmission
Baud rate: 1200, 2400, 4800, 9600 bps
Stop bits: 1 or 2 bits
Data length: 7 or 8 bits
Parity: even, odd, none
Remote control by controllers (RMT lamp lit)
Data output to the printer (manual, interval, help)
(PRINT lamp lit)
It is possible to set the interval with unit in 10 seconds from 10
seconds to 100 hours.
Remote release by LOCAL key (RMT lamp off)
(21) System reset
The units's settings are returned to the initial settings.