41
────────────────────────────────────────────────────
4.4 Setting the Number of Averaging
────────────────────────────────────────────────────
ENTER
Increasing
Decreasing
NOTE
Unit flashes
NOTE
Display:
Supplement
Key Key operation
PT・CT・SC
COMP1
COMP2
D/A
INTEG
PRINT
RANGE SEL
INTERFACE
Moves to each setting screen.
(2) Set the number of times for averaging.
1. Press the or key to change the averaging time in the desired column
when the displays on Display c are flashing. The setting range is 1 to 300.
2. Press the
ENTER
key, and the flashing will stop and the setting will be
decided.
Unless this operation is done, the internal setting cannot be changed.
At this time, if the averaging times are anything but "1", the AV (average)
indicator is lit.
To change the settings again, press the , , ,or key and flashing
indication appears.
The range of averaging time is 1 to 300. When the time for averaging is set
to 1, the AV indicator will turn off. When it is set to 000, pressing the
ENTER key changes the setting to 001. If it is set to more than 300, the
setting is made to 300.
(3) Move on to the measurement display.
Press the
AVE・FREQ
or
EXIT
key and the SHIFT indicator will turn off.
If the input of voltage, current or active power exceeds 130% of range
during averaging, the value is averaged and calculated as input of 130%f.s.
At that time, "o.r" is displayed. While an averaged value which includes an
out-of-range value is being displayed, the unit indication flashes to indicate
this fact.
Other keys operations in Averaging Setting mode
(SHIFT indicator is lit):