43
────────────────────────────────────────────────────
4.5 Setting the Frequency
────────────────────────────────────────────────────
ENTER
Setting frequency
source and range
Switching display
Display:
Supplement
Key Key operation
PT・CT・SC
COMP1
COMP2
D/A
INTEG
PRINT
RANGE SEL
INTERFACE
Moves to each setting screen.
NOTE
(2) Set the frequency.
1. To select the frequency source, press the or key while the indication
of V, A on display d is flashing.
2. To select the frequency range, press the key and the setting display (500
Hz, 100 kHz, or AUTO) on display d flashes. Press the or key to set
the range.
3. Pressing the or key moves flashing point.
4. Press the
ENTER
key, and the flashing will stop and the setting will be
decided.
Unless this operation is done, the internal setting cannot be changed.
If the settings need to be changed again, press either the , , , Setup
keys and the flashing points will reappear.
(3) Move onto the measurement display.
Press the
AVE・FREQ
or
EXIT
key and the SHIFT indicator will turn off.
Other keys operations in Frequency Setting mode
(SHIFT indicator is lit):
When the frequency range is set to "AUTO", and if the frequency exceeds
500 Hz, the range is set to 100 kHz, or if the frequency is less than 400 Hz,
the range is set to 500 Hz. If the range is not decided, "-----Hz" is
displayed.