45
────────────────────────────────────────────────────
4.6 Setting the D/A Output
────────────────────────────────────────────────────
ENTER
VAWVAvarPF
lp Hz Ah +Wh -Wh ±Wh
Display:
Supplement
Key Key operation
AVE・FREQ
PT・CT・SC
COMP1
COMP2
INTEG
PRINT
RANGE SEL
INTERFACE
Moves to each setting screen.
Figure 4 D/A Output of Phase Angle
Figure 3 D/A Output of Power Factor
0V
0
100%f.s.
Voltage, Current, Apparent power, Current peak
5V
Figure 1 D/A Output of Apparent Power
-5V
-100 % f.s.
0
100%f.s.
Active power, Reactive power
5V
0V
Figure 2 D/A Output of Reactive Power
-0
-0.5
1+0.5
+0
Power factor
-5V
5V
0V
LEAD LAG
-180°
+180°
LEAD LAG
Phase angle
0°
-5V
5V
0V
(2) Set the D/A output.
1. The measured data output from D/A is displayed as unit symbols on display
d. When the indication on display d is flashing, the units can be changed the
following way, by using the Setup keys.
2. Using the and keys switches the items.
3. Press the
ENTER
key, and the flashing will stop and the setting will be
decided.
Unless this operation is done, the internal setting cannot be changed.
If the settings need to be changed again, press either the , , , keys
and the flashing points will reappear.
(3) Move onto the measurement display.
Press the
D/A
or
EXIT
key and the SHIFT indicator will turn off.
Other keys operations in Averaging Setting mode (SHIFT indicator is lit):
For D/A output example, refer to Figure 1 to 6.