60
────────────────────────────────────────────────────
4.11 Displaying Comparator
────────────────────────────────────────────────────
COMP CH1 CH2 Normal display
0123456789OFF
CH1-1 V A W VA var PF
○
lp Hz Ah +Wh -Wh ±Wh TIME
NOTE
01OFF
Comparator item
Measurement value
Limit value
(2) Key operation
Key operations on the comparator display screen are as follow:
Pressing the
CH
or
TABLE
key in normal mode moves to Comparator display
screen.
When the
CH
key is pressed:
When the TABLE key is pressed:
Pressing the
c
key changes to the desired measurement item on Display c.
Pressing the
CH
or
TABLE
key displays comparator channel and table number.
Pressing the
a
,
b
,or
d
key displays comparator channel and table number for
one second, when the any measurement item is selected on Display c.
・ If the
CH
key and the
TABLE
key are pressed when all tables for Comparator
Ch1 and Ch2 set to OFF, "OFF" is displayed in display "d" for about 1
second, and then the previous display returns.
・ Tables set to OFF in comparator table settings are skipped when switching
table numbers with the
TABLE
key.
・ When the comparator is ON, comparator decisions are carried out according
to new table data immediately after switching table numbers with the
TABLE
key.
・ When the comparator item display "c" is displaying other items, comparator
information can be displayed by pressing the
CH
key and the
TABLE
key. In
addition, when the
a
or
b
key is pressed, comparator information is displayed
in display "c" for about 1 second, and then display of the measurement item
returns.
・ When the comparator is turned ON and normal display is set, information
for the currently set comparator item is displayed when the
CH
key or
TABLE
key is pressed. (The table is not switched; only the display is switched.)
Pressing the