A9
Zero Adjustment
Constant-current power supply
Voltmeter
SOURCE-H
SENSE-H SENSE-L
SOURCE-L
E
I
0
R
SEH
R
SOH
R
SOL
R
SEL
R
Short
I
E
=(
I
0
×
R
SEL
)+(
I
0
×
R
SEH
)
=(0×
R
SEL
)+(0×
R
SEH
)
=0 [V]
Figure. Conditions where a pseudo 0
Ω
is connected
To properly carry out the zero adjustment:
“Table. Connection Method” illustrates the correct connection method and the incorrect connection method.
The resistances in the gure show the wiring resistances, and these are ignored if they are less than several
Ω
respectively.
As shown in (a), when connecting respectively SENSE-H and SENSE-L, and SOURCE-H and SOURCE-L,
and then connecting the SENSE and the SOURCE with one path, there is no voltage potential difference
produced between SENSE-H and SENSE-L, and thus, a voltage of 0 V is applied. This connection method
performs the correct zero adjustment.
However, as illustrated (b), when connecting respectively SENSE-H and SOURCE-H, and SENSE-L and
SOURCE-L, and then connecting the Hi-side and the Lo-side with one path, there is a voltage potential
difference of
I
×
R
Short
between SENSE-H and SENSE-L. Thus, the connection method does not create
conditions where a pseudo resistance of 0
Ω
is connected, and thus, does not perform the correct zero
adjustment.
10
9
8
7
6
5
4
3
2
1
Appx. Ind.