92
Timing Chart
10 V
t1
t2
t6
t3 t5
t4
/START
(Input)
/TEST
(Output)
Test voltage
Measured
value display
/PASS
(Output)
/U.FAIL
/L.FAIL
(Output)
/ERR
(Output)
t7
Measurement end
When the TEST signal OFF
timing setting is SLOW
When measured
value-1 result is FAIL
When measured
value-2 result is PASS
Measurement-1 Measurement-2
Previous measured value
Previous result
Previous result
Previous result
OFF OFF ON
OFF ON OFF
OFF
0/0.0/0.00 M
Ω
Measured value-1 Measured value-2
Measurement-1 contact check result
Measurement,
judgment
(including
contact check)
Item Time
t1 START, STOP signal pulse width 5 ms
MIN.
t2 START, STOP signal detection time 5 ms
MAX.
*
t3 Response time (DELAY) AUTO, 5 ms to 999.9 s
t4 Measurement
time
Contact check: OFF 30 ms (FAST), 480 ms (SLOW)
Contact check: ON 80 ms (FAST), 480 ms (SLOW)
t5 Measurement
interval
Contact check: OFF 50 ms (FAST), 500 ms (SLOW)
Contact check: ON 100 ms (FAST), 500 ms (SLOW)
t6 Discharge time (time for the output voltage to drop
below 10 V)
20 ms
MAX.
(during pure resistance
measurement)
t7 Test duration (TIMER) 45 ms to 999.9 s
*
When the START signal is input after the test voltage is changed, a time of up to 500 ms is